Abstract
Background: Patients with facial paralysis require detailed clinical assessment and long-term follow-up to monitor facial function. The current standard of care for evaluating facial symmetry and movement uses validated clinician scoring tools such as the House-Brackmann facial paralysis score or the Electronic Clinician-Graded Facial Function Scale (eFACE). Existing tools are difficult to use in normal clinic workflows and do not provide real-time facial movement tracking, representing an unmet need. Therefore, we developed FaceADE, a novel iOS app leveraging native 3D image acquisition capabilities on the iPhone to rapidly quantify facial movement in patients with facial paralysis.
Objective: This study aimed to benchmark FaceADE against 2D image analysis and establish the feasibility of measuring oral commissure movement in patients with facial paralysis and healthy controls.
Methods: Patients were enrolled in a tertiary care clinic focused on facial paralysis treatment. Patients underwent image capture using the FaceADE app assisted by study team personnel. Measurements of lip commissure position and movement were obtained from 20 patients with facial paralysis and 10 healthy volunteers without facial paralysis. Measurements of lip movement and symmetry gathered from FaceADE were benchmarked against 2D measurements using open-source image analysis software (ImageJ).
Results: FaceADE measurements of lip commissure position and movement showed strong agreement with 2D measurements in both healthy volunteer and facial paralysis cohorts. The intraclass correlation coefficient was 0.96 (95% CI 0.90-0.98; P<.001) in the healthy volunteer cohort and 0.82 (95% CI 0.72-0.88; P<.001) in the facial paralysis cohort. Bland-Altman analysis found strong agreement between the 2 methods for these measurements. The 95% CI contained 97.5% (39/40) of data points in the healthy cohort and 91.2% (73/80) of data points in the facial paralysis cohort.
Conclusions: Our proposed mobile method of measuring clinically important lip commissure position and movement is feasible for use at the time of care delivery. In the future, this technology may be useful for a quantitative assessment of facial paralysis severity.
doi:10.2196/85965
Keywords
Introduction
Facial paralysis encompasses conditions that cause weakness or paralysis of the face. The incidence of facial paralysis in the United States was approximately 47.4 per 100,000 in 2022 [,]. Patients with facial paralysis report lower quality of life and severe restriction of emotional expression, eye closure, and communication [-]. Interventions may include medical treatment, physical therapy, and facial reanimation surgery for persistent and severe facial paralysis [-]. Currently, disease progression and recovery are tracked over time via patient-reported outcome measures and clinician-scored metrics based on physical examination [].
Facial paralysis is assessed by clinicians using several available scales, such as the Sunnybrook Facial Grading System, the Electronic Clinician-Graded Facial Function Scale (eFACE), and the House-Brackmann scale []. However, these scales may have significant interrater variability, and more detailed scales such as the eFACE may be difficult to complete in a busy clinical environment [,]. Thus, an automated and accurate measure of facial movement and symmetry could help standardize assessment and save valuable time during a clinical encounter.
Initial efforts to automate facial analysis have been made [-]. Existing methods allow for facial movement analyses from photographs or stored videos post hoc but do not offer real-time assessment and data output during the clinical encounter. Furthermore, they often rely on assumed average values, such as an iris size of 11.77 mm, to output common distance measures of facial movement [,].
We propose that Apple’s TrueDepth technology can be leveraged to provide real-time measurements of facial movement by creating an individualized 3D facial map [-]. For face recognition, the TrueDepth camera uses infrared (IR) projections to create a 3D map of the face. Facial features and movements can be measured despite variations in head tilt and skin tone [].
We developed an iOS app, FaceADE, to obtain immediate measurements of clinically important facial features using Apple’s TrueDepth camera. In this feasibility study, we demonstrate that this method can accurately measure oral commissure position and movement based on benchmark comparisons to 2D still-frame analysis in both healthy volunteers and patients with facial paralysis.
Methods
App Development
FaceADE is an iOS app developed in Swift (Apple Inc) to analyze facial features and quantify clinically relevant physical measurements using Apple’s TrueDepth camera []. This camera tracks projected IR dots to create a real-time 3D digital mesh of the face that retains accuracy in variable light conditions [,]. Apple’s ARKit uses this system to create a 1220-point face mesh with the coordinate system based relative to the position of the head []. The FaceLandmarks tool was used to draw curvilinear lines between the predetermined vector points that designate specific facial features of interest on the left and right hemiface []. Measurements of these curves were then imputed in the 3D space and then displayed within the app (). Facial measures include oral commissure position, mouth area, palpebral fissure height (eye height), and eye area. For this feasibility study, only oral commissure metrics were extracted and analyzed. The 16 facial measurements listed in were exported in a CSV file at 15 frames per second. Dynamic facial movement scores are reported as a percentage relative to the neutral expression, which can be set by the user at any point of the recording by pressing the “neutral expression” button. Patient demographic fields are shown in the lower left-hand corner above the “neutral expression” button. These fields allow for the labeling of patient data when they are exported after recording. In the lower right-hand corner, users can toggle on or off the face mesh, and to the right of the “neutral expression” button, there is a screenshot button to capture images while the video is recording. Measurements were captured at 15 frames per second on a tripod-stabilized iPad, and no video smoothing was applied. Measurements and vector coordinates of each frame of the video were collected to quantify movement during dynamic facial expressions. A walk-through video of FaceADE can be found in . FaceADE is available for download on GitHub [] and in the Apple App Store.

Facial Feature Measurements
Although the entire matrix of the face was tracked and recorded, the facial feature emphasized in this feasibility study was the lower lip commissure position (mm), defined as the distance from the right or left oral commissure to the facial midline along the lower lip vermillion border (). Oral commissure movement is one of the most clinically important features of facial movement and is the focus of many surgical procedures for facial paralysis [].
Ethical Considerations
Institutional review board approval was obtained (Stanford Institutional Review Board 79624), including an approved consent form and enrollment process. Patients aged 18 years and older of any gender were approved to be enrolled at the Stanford Facial Nerve Center. Eligible patients discussed study participation with a study team member in a private clinic room after their clinical encounter was complete. If they wished to participate in the study, patients subsequently completed the informed consent process. Study team members and clinical staff from the Facial Nerve Center served as controls. All patients and controls were assigned a sequential numerical identifier for data deidentification prior to analysis, and only study team members handled data for collection and analysis.
All images of individuals used in the manuscript are of study team members who provided explicit written consent for the use of their images and reviewed all figures prior to submission.
FaceADE videos were stored on encrypted and HIPAA (Health Insurance Portability and Accountability Act)-compliant iPads (iPad Pro; model number A2759; Apple Inc). Data were transferred from the iPads to a HIPAA-compliant server. Once videos were uploaded to the server, they were deleted from the iPads. All patient images were stored on the same secure server, and all data analysis was performed on a password-protected, encrypted, and HIPAA-compliant desktop computer.
Data Collection
Patient videos were obtained at the conclusion of the clinical encounter in a private examination room. After obtaining patient consent, a ruler was used to make a 1-cm mark with a surgical marker on the patient’s forehead. Patients were seated on a stool, and the tripod with the iPad mounted was placed in front of them. Patients were asked to look directly into the camera during recording. Patients were read standardized face examination prompts () and asked to rest their face between each prompt. The video was then uploaded to a HIPAA-compliant server. Videos were used to obtain still frames of the patient’s neutral and full smile expressions selected from a representative video frame of the face after the patient received the face examination prompt from the recording (ie, “please give a big smile”). The still frames were opened in ImageJ and scaled using the 1-cm mark on the skin. The segmented line tool in ImageJ was then used to measure right and left commissure distance along the lower lip vermillion border. To avoid bias in ImageJ measurements and analysis, the videos had the face mesh turned off, and ImageJ data were entered into REDCap (Research Electronic Data Capture; Vanderbilt University) before FaceADE data were recorded. A study flowchart is shown in .
Analysis
To measure the overall agreement between 2D analysis and FaceADE, the measurements from both facial sides and expressions were pooled, and the intraclass correlation coefficient was calculated using a 2-way random-effects model where both patients and the 2 methods were considered to be randomly selected, and absolute agreement was used instead of the consistency of the trend. The reliability estimate of a single rating was generated. To estimate the potential bias between methods, we performed separate Bland-Altman analyses for each cohort. The bias was defined as the difference between every pair of FaceADE and ImageJ measurements. For each cohort, we used a linear mixed model to account for the measurement of both sides of the face from the same patient. The models included the bias as the outcome measure, facial side and expression as fixed effects, and patient as a random effect with random intercept. We calculated the total variance by adding the within- and between-subject variation generated from the model. The 95% limits of agreement were computed as the bias ±1.96√total variance. Standard Bland-Altman plots were created, with the bias plotted against the average of FaceADE and ImageJ measurements. A horizontal line was drawn at the mean difference and 95% limits of agreement. The scatter dots outside the 95% limits of agreement were identified as possible outliers. All analyses were conducted using RStudio (version 2026.01.2+418; Posit PBC).
Results
Our healthy volunteer cohort was composed of 10 study team members, with 6 (60%) being female; the facial paralysis cohort included 20 patients, with 9 (45%) being female. The average age at paralysis onset was 45.5 (SD 21.1) years. The average duration of the videos was 78 (SD 27.5) seconds ().
| Values | ||||
| Facial paralysis (n=20) | Healthy volunteers (n=10) | |||
| Age at facial paralysis onset (years), mean (SD) | 45.5 (21.1) | —a | ||
| Age at consent (years), mean (SD) | 53.4 (16.5) | 30.1 (7.8) | ||
| Sex, n (%) | ||||
| Female | 9 (45) | 6 (60) | ||
| Male | 11 (55) | 4 (40) | ||
| Race, n (%) | ||||
| Asian | 7 (35) | 4 (40) | ||
| White | 7 (35) | 5 (50) | ||
| Other | 6 (30) | 1 (10) | ||
| Ethnicity, n (%) | ||||
| Hispanic or Latino | 6 (30) | 2 (20) | ||
| Non-Hispanic or Latino | 14 (70) | 8 (80) | ||
| Reason for FP, n (%) | ||||
| Bell\'s palsy | 8 (40) | — | ||
| Ramsay Hunt syndrome | 3 (15) | — | ||
| Iatrogenic | 2 (10) | — | ||
| Malignancy | 6 (30) | — | ||
| Right arteriovenous malformation | 1 (5) | — | ||
| Laterality of facial injury, n (%) | ||||
| Left | 9 (45) | — | ||
| Right | 11 (55) | — | ||
aNot applicable.
As a video- and IR-based method, FaceADE tracked facial movements over time. A healthy volunteer performed a small smile and a full smile with dental show, simulated a small asymmetrical smile, and simulated a larger asymmetrical smile with dental show (A). In an asymmetrical smile, FaceADE revealed greater movement of the right commissure and tracked the distortion of the left commissure that resulted from unilateral movement on one side. Analysis of commissure movement during a smile in a patient with facial paralysis showed similar trends while tracking oral commissure movement throughout the facial expression (B). A full smile with dental show resulted in greater commissure asymmetries than small smile expressions, and subtle changes in both commissures could be observed over the course of the movement.
To benchmark oral commissure movement tracking, we compared oral commissure movements calculated by FaceADE to those calculated using standard 2D analysis of still frames in ImageJ (). Among healthy volunteers, there was concordance between the left and right hemiface in both neutral expression (face in repose) and large smile expressions with dental display (A; intraclass correlation coefficient=0.96, 95% CI 0.90-0.98; P<.001). The SEs for commissure measurements in the healthy volunteer cohort during a neutral expression on the left and right sides with FaceADE and ImageJ were 0.45, 0.49, 0.72, and 0.70, respectively. The SEs for commissure measurements in the healthy volunteer cohort during a full smile were 0.72, 0.72, 0.77, and 0.62, respectively. Similarly, there was strong alignment between FaceADE and ImageJ in the facial paralysis cohort in both neutral and full smile expressions (B). The intraclass correlation coefficient within the facial paralysis cohort showed good reliability at 0.82 (95% CI 0.72-0.88; P<.001). The SEs for commissure measurements in the facial paralysis cohort during a neutral expression on the unaffected and affected sides with FaceADE and ImageJ were 0.62, 0.79, 0.85, and 1.00, respectively. The SEs for commissure measurements in the facial paralysis cohort during a full smile were 0.91, 1.06, 1.06, and 1.01, respectively. Bland-Altman analysis () revealed a mean bias of 1.13 mm in the healthy cohort and 0.45 mm in the facial paralysis cohort, indicating that FaceADE measurements were systematically larger than ImageJ measurements. The 95% limits of agreement were −2.13 to 4.38 mm in the healthy cohort and −5.97 to 6.87 mm in the facial paralysis cohort. In total, 97.5% (39/40) of the measurements fell within these limits for the healthy cohort, and 91.2% (73/80) of the measurements fell within these limits for the facial paralysis cohort. Disagreement between FaceADE and ImageJ oral commissure measurement was examined using the Fitzpatrick skin type test (), and the median disagreement for all skin types was found to be less than 5 mm.


Discussion
Principal Findings
Our findings highlight the feasibility of using IR- and video-based imaging to measure oral commissure symmetry and movement in healthy controls and patients with facial paralysis. In this pilot study, we aimed to benchmark FaceADE’s output against gold-standard 2D image analysis and test the feasibility of using this app during normal clinic workflow.
FaceADE has several advantages compared to existing options. Previously published approaches require a measurement scale to be set using an assumed average value, such as a presupposed iris diameter of 11.77 mm or interpupillary distance of 63 mm. FaceADE’s 3D mesh overlay is individualized to a patient’s face and rapidly created using native iOS software and device hardware []. Using both video and IR inputs, FaceADE extracts important temporal data such as the velocity of a facial expression and may be able to quantify difficulty maintaining a facial expression over time. Overall, video-based analysis may provide a more complete assessment of facial movement in comparison to 2D analysis of selected still frames. FaceADE captures the entire facial expression attempt, including maximum movement, average movement, and expression duration. The high intraclass correlation coefficient between FaceADE and gold-standard distance measures is encouraging.
FaceADE is designed to fit into existing clinical workflows. The videos take under 2 minutes to record, and the app is run on the existing iOS platform without the need for additional hardware. The method does not require key frame selection, manual adjustment of key points at anatomical landmarks, or other post hoc modification of images. It also allows for the capture and local storage of videos and still-frame images. With these data, clinicians may be able to use this technology to track patient facial function and recovery over time. A future patient-facing version may allow patients to collect data and share them with their health care team if integrated with a HIPAA-compliant storage platform and electronic medical record.
This feasibility study focused only on oral commissure measurements; therefore, the accuracy of tracking other facial features will need to be validated in future work. Of note, measurements obtained from ImageJ are 2D, whereas FaceADE measures movement in the x-, y-, and z-axes over time. This may at least partially account for the higher distance output by FaceADE than by ImageJ. The ARKit technology that FaceADE leverages is developed by Apple Inc []. The accuracy of this method is currently limited to Apple’s proprietary technology and algorithms. While we observed a strong intraclass correlation, the limits of agreement in the Bland-Altman analysis were rather wide, suggesting the need for deeper subgroup analysis to identify weaknesses within the ARKit used in FaceADE. Disease severity or other patient factors may contribute to a higher error rate, and this will be a focus of future study.
Additionally, the sample size in this feasibility study was too small to evaluate accuracy and full clinical utility. Critically, we must assess the reliability and accuracy of FaceADE across patients with different disease severities (as judged by clinician experts using validated metrics such as the Sunnybrook Facial Grading System or eFACE scale) and facial anatomical characteristics, test-retest scenarios, and other common clinical variables. This will be the focus of a future validation study.
While we focused on oral commissure movement for this study as a proof of concept, FaceADE collects data from 1220 vectors on the face. In theory, this permits an exhaustive analysis of facial movement that includes all major facial expressions routinely captured during clinical evaluation of facial paralysis and will provide deeper insights into the diagnosis and treatment of this condition.
Conclusions
FaceADE offers a feasible and mobile method to use 3D landmarks to track oral commissure movement and symmetry in healthy controls and patients with facial paralysis. This framework could be expanded to provide clinicians with an easy-to-use system to objectively record facial paralysis severity. Video analysis and 3D modeling allow for potentially more accurate and detailed measurements compared to existing technologies.
Acknowledgments
The authors would like to thank Mr. Yifei Ma from the Stanford Otolaryngology – Head & Neck Surgery Department for assisting with statistical analysis. This work was presented at the 15th International Facial Nerve Symposium in Madrid, Spain, in September 2025. ChatGPT (version GPT-5; OpenAI) was used as an iterative coding support tool during app development. All study ideas and app features are original to the authors; artificial intelligence assistance was limited to debugging and troubleshooting errors encountered during feature development.
Funding
This work was funded by the Stanford University School of Medicine Otolaryngology – Head & Neck Surgery Department and Stanford Institute for Human-Centered AI.
Authors' Contributions
MP, AN, KG, AS, and JPP contributed to the conception and design of this work. MP, AN, KG, AAM, AS, and JPP contributed to data acquisition. MP, AS, AAM, and JPP contributed to statistical analysis. MP, AN, and AS contributed to drafting the manuscript. MP, AN, AAM, AS, and JPP contributed to analysis and interpretation of the data. MP, AN, KG, AAM, AS, and JPP contributed to critical revision of the manuscript for important intellectual content.
Conflicts of Interest
None declared.
Video walk-through of the FaceADE app and its features.
MP4 File (MP4 Video), 190322 KBCollected FaceADE features.
DOCX File , 445 KBStandardized prompts used for facial analysis in patients with facial paralysis.
PNG File , 58 KBWorkflow diagram of the process of data collection, encryption, and analysis.
PNG File , 204 KBBland-Altman analysis of agreement between FaceADE and ImageJ oral commissure measurements.
DOCX File , 67 KBDistribution of disagreement between FaceADE and ImageJ oral commissure measurements using the Fitzpatrick skin type test.
DOCX File , 76 KBReferences
- Wei EX, Green A, Pepper JP, Akkina SR. Increasing incidence of facial nerve disorders in the United States from 2007 to 2022. Laryngoscope. Jun 2025;135(6):2008-2013. [CrossRef] [Medline]
- Ragaban A, Alsharif L, Alshaikh NA, Jafar RJ, Hemeq Z, Khan MA, et al. Prevalence, etiology, risk factors, and complications of facial nerve palsy at King Abdulaziz Medical City: a multicenter study. Cureus. Feb 1, 2024;16(2):e53403. [FREE Full text] [CrossRef] [Medline]
- Nellis JC, Ishii M, Byrne PJ, Boahene KD, Dey JK, Ishii LE. Association among facial paralysis, depression, and quality of life in facial plastic surgery patients. JAMA Facial Plast Surg. May 01, 2017;19(3):190-196. [FREE Full text] [CrossRef] [Medline]
- de Swart BJ, Verheij JC, Beurskens CH. Problems with eating and drinking in patients with unilateral peripheral facial paralysis. Dysphagia. 2003;18(4):267-273. [CrossRef] [Medline]
- Kim JH, Fisher LM, Reder L, Hapner ER, Pepper JP. Speech and communicative participation in patients with facial paralysis. JAMA Otolaryngol Head Neck Surg. Aug 01, 2018;144(8):686-693. [FREE Full text] [CrossRef] [Medline]
- Hohman MH, Mazzoni T. Facial nerve palsy. In: StatPearls. Treasure Island, FL. StatPearls Publishing; 2025.
- Ravikumar A, Singh P, Batish VK. Facial palsy - treatment options. Med J Armed Forces India. Jan 1999;55(1):41-44. [FREE Full text] [CrossRef] [Medline]
- Volk GF, Klingner C, Finkensieper M, Witte OW, Guntinas-Lichius O. Prognostication of recovery time after acute peripheral facial palsy: a prospective cohort study. BMJ Open. Jun 20, 2013;3(6):e003007. [FREE Full text] [CrossRef] [Medline]
- Finsterer J. Management of peripheral facial nerve palsy. Eur Arch Otorhinolaryngol. Jul 2008;265(7):743-752. [FREE Full text] [CrossRef] [Medline]
- Ho AL, Scott AM, Klassen AF, Cano SJ, Pusic AL, Van Laeken N. Measuring quality of life and patient satisfaction in facial paralysis patients: a systematic review of patient-reported outcome measures. Plast Reconstr Surg. Jul 2012;130(1):91-99. [CrossRef] [Medline]
- Tedeschi R, Donati D, Giorgi F. Beyond the smile: a systematic review of diagnostic tools for peripheral facial paralysis. Acta Neurol Belg. Dec 2024;124(6):1805-1822. [CrossRef] [Medline]
- Mengi E, Orhan Kara C, Necdet Ardıç F, Topuz B, Metin U, Alptürk U, et al. Comparison of the reliability of the House- Brackmann, Facial Nerve Grading System 2.0, and Sunnybrook Facial Grading System for the evaluation of patients with peripheral facial paralysis. J Int Adv Otol. Jan 2024;20(1):14-18. [FREE Full text] [CrossRef] [Medline]
- Jeong J, Lee JM, Cho YS, Kim J. Inter-rater discrepancy of the House-Brackmann facial nerve grading system. Clin Otolaryngol. Nov 2022;47(6):680-683. [CrossRef] [Medline]
- Baltrusaitis T, Zadeh A, Lim YC, Morency LP. OpenFace 2.0: facial behavior analysis toolkit. In: Proceedings of the 13th IEEE International Conference on Automatic Face & Gesture Recognition. 2018. Presented at: FG 2018; May 15-19, 2018; Xi'an, China. [CrossRef]
- Guarin DL, Dusseldorp J, Hadlock TA, Jowett N. A machine learning approach for automated facial measurements in facial palsy. JAMA Facial Plast Surg. Jul 01, 2018;20(4):335-337. [CrossRef] [Medline]
- jeffheaton/dynaface. GitHub. URL: https://github.com/jeffheaton/dynaface [accessed 2025-08-13]
- TadasBaltrusaitis/OpenFace. GitHub. URL: https://github.com/TadasBaltrusaitis/OpenFace/blob/master/README.md [accessed 2025-08-13]
- Banks CA, Bhama PK, Park J, Hadlock CR, Hadlock TA. Clinician-graded electronic facial paralysis assessment: the eFACE. Plast Reconstr Surg. Aug 2015;136(2):223e-230e. [CrossRef] [Medline]
- Gaudin RA, Robinson M, Banks CA, Baiungo J, Jowett N, Hadlock TA. Emerging vs time-tested methods of facial grading among patients with facial paralysis. JAMA Facial Plast Surg. Jul 01, 2016;18(4):251-257. [CrossRef] [Medline]
- Özsoy Ö, Özsoy U, Yıldırım Y, Yüksek HE. Evaluating facial paralysis: a comparative study of 2D Emotrics and 3D scanning technologies. J Plast Reconstr Aesthet Surg. Dec 2025;111:53-57. [CrossRef] [Medline]
- Hasebe K, Kojima T, Okanoue Y, Yuki R, Yamamoto H, Otsuki S, et al. Novel evaluation method for facial nerve palsy using 3D facial recognition system in iPhone. Auris Nasus Larynx. Jun 2024;51(3):460-464. [FREE Full text] [CrossRef] [Medline]
- Taeger J, Bischoff S, Hagen R, Rak K. Utilization of smartphone depth mapping cameras for app-based grading of facial movement disorders: development and feasibility study. JMIR Mhealth Uhealth. Jan 26, 2021;9(1):e19346. [FREE Full text] [CrossRef] [Medline]
- Müller SL, Pfister P, Menzi N, Muller L, Klein HJ, Schweizer R, et al. Artificial intelligence in facial palsy treatment: a systematic review and recommendations. Plast Reconstr Surg. Sep 01, 2025;156(3):477-490. [CrossRef] [Medline]
- ARFaceAnchor. Apple Developer Documentation. URL: https://developer.apple.com/documentation/arkit/arfaceanchor [accessed 2025-08-13]
- Swift. Apple Developer Documentation. URL: https://developer.apple.com/documentation/swift/ [accessed 2025-08-12]
- ARKit face mesh vertex tool. FaceLandmarks. URL: https://facelandmarks.com/ [accessed 2025-08-12]
- About Face ID advanced technology. Apple Support. URL: https://support.apple.com/en-us/102381 [accessed 2025-08-12]
- ARKit in iOS. Apple Developer Documentation. URL: https://developer.apple.com/documentation/arkit/arkit-in-ios [accessed 2025-08-12]
- ktheguo/faceADE. GitHub. URL: https://github.com/ktheguo/faceADE [accessed 2026-06-30]
- Kim L, Byrne PJ. Controversies in contemporary facial reanimation. Facial Plast Surg Clin North Am. Aug 2016;24(3):275-297. [CrossRef] [Medline]
Abbreviations
| eFACE: Electronic Clinician-Graded Facial Function Scale |
| HIPAA: Health Insurance Portability and Accountability Act |
| IR: infrared |
| REDCap: Research Electronic Data Capture |
Edited by J Sarvestan; submitted 15.Oct.2025; peer-reviewed by AK Vadathya, S Mohan, JBW Weiss; comments to author 26.Feb.2026; revised version received 14.Jun.2026; accepted 15.Jun.2026; published 14.Jul.2026.
Copyright©Marcelina Puc, Katherine Guo, Anthony Newman, Alexandra Antoinette Myers, Amar Sheth, Jon-Paul Pepper. Originally published in JMIR Formative Research (https://formative.jmir.org), 14.Jul.2026.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Formative Research, is properly cited. The complete bibliographic information, a link to the original publication on https://formative.jmir.org, as well as this copyright and license information must be included.