Research Letter
Abstract
Among newly postpartum people, the novel Just Breathe guided breathing device showed high user satisfaction scores and self-perceived reductions in stress, although the Cohen d effect on heart rate variability and stress or anxiety symptoms as measured by validated psychometric scales was minimal.
JMIR Form Res 2026;10:e85321doi:10.2196/85321
Keywords
Introduction
Over 1 in 5 birthing people are diagnosed with perinatal mood or anxiety disorders, and many who do not meet diagnostic thresholds for these conditions experience stress or anxiety that negatively impacts infant bonding or decrease breastfeeding rates [-]. Mindful breathing can improve mood and stress symptoms and increase heart rate variability (HRV) []. Defined as the beat-to-beat variability in heart rate, HRV is inversely associated with symptoms of depression and stress and can be assessed using wrist monitors []. A novel device, Just Breathe, incorporates airflow-guided cues that optimize adherence to evidence-based mindful breathing patterns shown to reduce stress symptoms []. Just Breathe is lifted to the user’s slightly opened lips, and it directs gentle airflow into the mouth to prompt inhalation and on the upper lip to cue exhalation over a 3-minute cycle (Figure S1 in ); it is yet to be examined in clinical settings. Our primary aim was to assess Just Breathe’s feasibility, acceptability, and perceived effectiveness among newly postpartum people. Our secondary, exploratory aim was to explore its effects on HRV, anxiety symptoms, and perceived stress.
Methods
Participants were recruited at a tertiary care women’s hospital. Eligible patients were English-speaking, ≥18 years old, and admitted to the postpartum unit during their childbirth hospitalization. Patients were excluded if incarcerated or if their infant required neonatal intensive care unit (NICU) admission. After providing consent, participants completed demographic surveys and the Perceived Stress Scale-4 (PSS-4) [] and Generalized Anxiety Disorder-7 (GAD-7) [] scale to assess for symptoms of stress and anxiety, respectively. A research staff member applied a Kairos Biostrap wrist monitor, a commercially available high-grade wearable that uses infrared photoplethysmography to assess HRV, measured in root mean square of successive differences. The wrist strap monitor was connected to a smartphone app, Vital Science, which summarizes HRV during the 3-minute cycle. After the participant rested in her hospital bed or chair for 5 minutes without interruption, one Vital Science cycle was conducted before Just Breathe use, and one Vital Science cycle was conducted while the participant used Just Breathe for 3 minutes. Participants were provided a Just Breathe device and encouraged to use it at least 3 times for up to 24 hours. The day after consent, a research staff member repeated the same procedures to assess HRV.
Participants then completed PSS-4, GAD-7, system usability scale (SUS; assesses patient experience with ease of use), Client Satisfaction Questionnaire (CSQ; assesses patient satisfaction with device), and the Decisional Regret Scale (assesses regret after making a health-related decision). To preserve validity and reliability of the surveys while ensuring they pertained to Just Breathe, SUS and CSQ were minimally modified by replacing the words “system” and “service,” respectively, with the words “Just Breathe device.” Descriptive statistics (means, SD, and percentages) compared initial and final HRV assessments and psychometric survey results. For HRV, GAD-7, and PSS-4, effect sizes for pre-post changes were calculated using paired Cohen d. Hypothesis testing was not performed due to the sample size. This study was approved by the institutional review board at Women and Infant Hospital of Rhode Island before study initiation (WIH 24-0049; approved on October 14, 2024). Informed consent was obtained, all data were de-identified, and participants could opt out per their preference. Those who completed the study received a US $5 gift card.
Results
Among 46 consented participants, 40 completed all survey measures and were included. Participants were racially, ethnically, and socioeconomically diverse: 48% (19/40) were non-White or Hispanic and 40% (16/40) had federal health insurance. Among the 40 study participants, 83% (n=33) were mostly or very satisfied with Just Breathe, and most (n=30, 75%) believed the device met their needs. In terms of perceived effectiveness, 88% (n=35) felt that Just Breathe helped improve their symptoms of stress or anxiety, and 85% (n=34) would recommend the device to a friend who was pregnant or newly delivered ().
Regarding exploratory analyses on psychometric outcomes (), although Just Breathe use was associated with a 25% increase in mean HRV, Cohen d analyses suggested a minimal effect of Just Breathe on HRV, GAD-7 scores, and PSS-4 scores.
| Question | Value, n (%) | ||
| How satisfied are you with the amount of information Just Breathe provided? | |||
| Dissatisfied/indifferent | 7 (17) | ||
| Mostly/very satisfied | 33 (83) | ||
| To what extent has Just Breathe met your needs? | |||
| Only a few/no needs met | 10 (25) | ||
| Most/all needs met | 30 (75) | ||
| Did Just Breathe help you to deal more effectively with symptoms of stress or anxiety? Answer: Yes | 35 (88) | ||
| If a friend were pregnant or newly delivered, would you recommend Just Breathe? Answer: Yes | 34 (85) | ||
| If you were able to, would you use Just Breathe again? Answer: Yes | 29 (73) | ||
| Measure | Before Just Breathe, mean (SD) | After Just Breathe, mean (SD) | Change pre/post Just Breathea (%) | Paired Cohen d |
| Heart rate variability (ms) | 36.0 (19) | 39 (18) | 26 | 0.21 |
| Perceived Stress Score-4 | 7.9 (2.3) | 7.4 (2.5) | –8 | –0.18 |
| Generalized Anxiety Disorder-7 | 4.3 (3.7) | 4.7 (3.8) | 0.4 | 0.21 |
aFor the pre/post change, no absolute value could be calculated, as only the change in mean (SD) was evaluated.
Discussion
Among newly postpartum people, Just Breathe was feasible and acceptable as a stress- and anxiety-reduction intervention, with high satisfaction ratings and strong user perception of benefit. Our exploratory analyses showed a minimal effect on HRV or psychometric outcomes after using Just Breathe for less than 24 hours. This could be because the device had no measurable effect or due to study methodology. Specifically, the study occurred in the early postpartum period over a 24-hour time period. Postpartum hospitalization is characterized by nursing- and infant-derived sleep interruption, delivery-related pain, and hormonal fluctuations that may increase stress and anxiety, thereby reducing Just Breathe’s effect. Prior studies that demonstrated an association between mindful breathing and improved postpartum mood or stress symptoms required participants to use mindful breathing for up to 8 weeks [,]. Thus, the lack of a strong signal in our psychometric outcomes may be due to a short study period within a backdrop of profound physiologic and psychologic adjustments. This study’s strengths include using both physiological and psychological measures, high participant retention, and inclusion of a racially, ethnically, and socioeconomically diverse study population. Limitations include the single-arm design, small sample size, short study duration, and exclusion of non-English speaking people and those whose infants were in the NICU, who are at higher rates of having stress or anxiety symptoms. Just Breathe is a novel device that is acceptable and viewed as effective. We did not assess the frequency that participants used Just Breathe between enrollment and study conclusion, which can be considered a major limitation that precludes the analysis of a dose-response relationship, thereby limiting the conclusions that could be drawn on Just Breathe’s effect. In conclusion, a randomized controlled trial using Just Breathe as a study intervention for weeks, if not months, is needed to more accurately determine whether this promising device can improve perinatal physiologic regulation and symptoms of stress and anxiety after childbirth.
Acknowledgments
Artificial intelligence was not used in any capacity to generate or edit this manuscript.
Funding
AKL is supported by the Eunice Kennedy Shriver National Institute of Child Health and Human Development (K23HD103961 and R01HD116695). MHB is supported by the National Heart, Lung, and Blood Institute (R01HL157288 and R01HL172869). These funding sources had no role in study design, collection, analysis, or interpretation of the data or in the decision to submit the article for publication.
Conflicts of Interest
AKL served as a paid consultant for Sonio AI in 2024. The company that created Just Breathe provided 10 free Just Breathe devices, one Kairos Biostrap wrist monitor, and access to the Vital Science smartphone app for this study but was not otherwise engaged with the research. Specifically, the design, analysis, or reporting of study findings all occurred independently with the investigative team without any input from any technology company.
Authors' Contributions
Conceptualization: AKL (lead), NKA (equal), MHB (supporting)
Data curation: SDN (lead), JLR (equal)
Formal analysis: TLJ (lead)
Funding acquisition: n/a
Investigation: SKS (lead), LS (equal), KB (supporting), JRL (supporting)
Methodology: AKL (lead)
Project administration: CFW (lead), SDN (equal)
Resources: TLJ (lead), CFW (equal)
Supervision: AKL (lead), NKA (equal)
Validation: MHB (lead), TLJ (supporting)
Visualization: TLJ (lead), KB (supporting)
Writing – original draft: SKS (lead), AKL (equal)
Writing – review & editing: NKA (lead), MHB (equal), SDN (supporting), LS (supporting), TLJ (supporting), JLR (supporting), CFW (supporting), KB (supporting)
Diagram of the Just Breathe device mouthpiece and airflow haptic system.
DOCX File , 293 KBReferences
- Ko JY, Rockhill KM, Tong VT, Morrow B, Farr SL. Trends in postpartum depressive symptoms - 27 states, 2004, 2008, and 2012. MMWR Morb Mortal Wkly Rep. Feb 17, 2017;66(6):153-158. [FREE Full text] [CrossRef] [Medline]
- American COO. ACOG committee opinion no. 757: screening for perinatal depression. Obstet Gynecol. Nov 2018;132(5):e208-e212. [CrossRef] [Medline]
- Aris-Meijer J, Bockting C, Stolk R, Verbeek T, Beijers C, van Pampus M, et al. What if pregnancy is not seventh heaven? The influence of specific life events during pregnancy and delivery on the transition of antenatal into postpartum anxiety and depression. Int J Environ Res Public Health. Aug 09, 2019;16(16):2851. [FREE Full text] [CrossRef] [Medline]
- Kim H, Cheon E, Bai D, Lee YH, Koo B. Stress and heart rate variability: a meta-analysis and review of the literature. Psychiatry Investig. Mar 2018;15(3):235-245. [FREE Full text] [CrossRef] [Medline]
- Li K, Cardoso C, Moctezuma-Ramirez A, Elgalad A, Perin E. Heart rate variability measurement through a smart wearable device: another breakthrough for personal health monitoring? Int J Environ Res Public Health. Dec 06, 2023;20(24):7146. [FREE Full text] [CrossRef] [Medline]
- Chaitanya S, Datta A, Bhandari B, Sharma VK. Effect of resonance breathing on heart rate variability and cognitive functions in young adults: a randomised controlled study. Cureus. Feb 2022;14(2):e22187. [FREE Full text] [CrossRef] [Medline]
- Cohen S, Kamarck T, Mermelstein R. A global measure of perceived stress. Journal of Health and Social Behavior. Dec 1983;24(4):385. [CrossRef]
- Spitzer RL, Kroenke K, Williams JBW, Löwe B. A brief measure for assessing generalized anxiety disorder: the GAD-7. Arch Intern Med. May 22, 2006;166(10):1092-1097. [CrossRef] [Medline]
- Sheydaei H, Ghasemzadeh A, Lashkari A, Kajani PG. The effectiveness of mindfulness training on reducing the symptoms of postpartum depression. Electron Physician. Jul 2017;9(7):4753-4758. [FREE Full text] [CrossRef] [Medline]
- Luberto CM, Park ER, Goodman JH. Postpartum outcomes and formal mindfulness practice in mindfulness-based cognitive therapy for perinatal women. Mindfulness (NY). Jun 2018;9(3):850-859. [FREE Full text] [CrossRef] [Medline]
Abbreviations
| CSQ: Client Satisfaction Questionnaire |
| GAD-7: Generalized Anxiety Disorder-7 |
| HRV: heart rate variability |
| NICU: neonatal intensive care unit |
| PSS-4: Perceived Stress Scale-4 |
| SUS: system usability scale |
Edited by J Sarvestan; submitted 05.Oct.2025; peer-reviewed by J Guo; comments to author 31.Dec.2025; revised version received 22.Jan.2026; accepted 23.Jan.2026; published 27.Feb.2026.
Copyright©Sarai K Sales, Stephanie D Nunez, Lindsay Spratt, Tracy L Jackson, Janet L Rousseau, Crystal F Ware, Kristopher L Britton, Margaret H Bublitz, Nina K Ayala, Adam K Lewkowitz. Originally published in JMIR Formative Research (https://formative.jmir.org), 27.Feb.2026.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Formative Research, is properly cited. The complete bibliographic information, a link to the original publication on https://formative.jmir.org, as well as this copyright and license information must be included.