Evaluation of a Digital Handheld Hydrogen Breath Monitor to Diagnose Lactose Malabsorption: Interventional Crossover Study

Evaluation of a Digital Handheld Hydrogen Breath Monitor to Diagnose Lactose Malabsorption: Interventional Crossover Study

Evaluation of a Digital Handheld Hydrogen Breath Monitor to Diagnose Lactose Malabsorption: Interventional Crossover Study

Original Paper

1Johns Hopkins Medicine, Baltimore, MD, United States

2Electronics Program, Penn Foster College, Scottsdale, AZ, United States

3Duke Early Phase Clinical Research Unit, Durham, NC, United States

Corresponding Author:

Simon C Mathews, MD

Johns Hopkins Medicine

1800 Orleans St

Baltimore, MD, 21287

United States

Phone: 1 410 955 5000

Email: smathe14@jhmi.edu


Background: Lactose malabsorption is a common condition that affects a broad segment of the population. Clinical diagnosis based on symptom recall can be unreliable and conventional testing can be inconvenient, requiring expensive laboratory-based equipment and conduction of the testing in a clinical setting.

Objective: The aim of this study is to assess the performance of a digital handheld hydrogen breath monitor (GIMate) in diagnosing lactose malabsorption compared to a US Food and Drug Administration (FDA)–cleared device (H2 Check) for the same indication.

Methods: An interventional crossover study was performed in adult participants with a prior confirmed diagnosis of lactose malabsorption or a suspected history of lactose intolerance.

Results: A total of 31 participants (mean age 33.9 years) were enrolled in the study. There was 100% positive percent agreement and 100% negative percent agreement between the GIMate monitor and the H2 Check. Correlation between gastrointestinal symptoms and hydrogen values was positive at 0.82 (P<.001).

Conclusions: The digital handheld GIMate breath monitor achieved equivalent diagnostic performance to that of an FDA-cleared device in the diagnosis of lactose malabsorption.

Trial Registration: ClinicalTrials.gov NCT04754724; https://clinicaltrials.gov/ct2/show/NCT04754724

JMIR Form Res 2021;5(10):e33009

doi:10.2196/33009

Keywords



Lactose malabsorption is a common condition due to lactase deficiency; for many, it results in gastrointestinal symptoms, which is termed lactose intolerance [1]. Lactase is an enzyme occurring in the intestinal mucosa that hydrolyzes lactose into its constituent parts, galactose and glucose [2]. The enzyme is normally present in neonates; however, for a majority of individuals worldwide, there is an inherited and irreversible reduction in enzyme activity as individuals age [3]. Secondary lactose malabsorption can also occur when there is injury to the intestinal mucosa from a reversible condition, such as infection [4]. Not all individuals with lactose malabsorption will experience bothersome gastrointestinal symptoms (ie, lactose intolerance). Those with lactose intolerance are often diagnosed clinically—that is, their response is observed to lactose challenges that come through dietary exposure, followed by a trial of avoidance of lactose-containing products [5]. However, self-reported intolerance can often be unreliable [6,7], and formal diagnosis is still helpful in many cases.

Conventional testing is conducted through a hydrogen breath test in the ambulatory clinical setting. Individuals usually present to the testing site fasting, and a baseline breath reading is obtained. They are subsequently challenged with a lactose-containing solution, with follow-up hydrogen readings obtained on an hourly basis for 3 hours [8].

The mechanism of hydrogen detection is based on undigested lactose in the colon being fermented by bacteria and resulting in the production of hydrogen, which is then partially absorbed into the bloodstream and ultimately exhaled by the lungs via the pulmonary circulation and gas exchange. Direct lactase activity can also be measured on tissue obtained through jejunal biopsy via endoscopy. This approach, however, is more invasive, costly, and potentially less reliable given issues relating to sampling bias [5].

Current methods for hydrogen breath testing can be costly for physicians to purchase and inconvenient for patients because they must take time out of their day to go to the testing site for several hours. In addition, given that most conventional breath-testing equipment is reusable, infectious contamination can occur through aerosolized breath contents. As a result, the validation of an alternative breath test that is portable, handheld, and disposable presents an opportunity to improve the value, safety, and experience associated with hydrogen breath testing.

In this study, our primary aim was to compare the performance of the GIMate (Vivante Health), a novel digital handheld hydrogen breath monitor, to that of the H2 Check (Micro Medical Limited), a device cleared by the US Food and Drug Administration (FDA) for the diagnosis of lactose malabsorption.


Study Design and Procedure

The study was an interventional crossover design, with all participants receiving both diagnostic interventions. The order of which intervention was received first was alternately assigned at random (Multimedia Appendix 1). The study was conducted at Duke University and was reviewed and approved by the Duke University Institutional Review Board. Upon screening as eligible for the study, participants were provided with best practice pre–breath testing guidance [9], including dietary restrictions and fasting overnight. Baseline breath hydrogen measurements were performed the following day using the GIMate and H2 Check. This was followed by ingestion of a 25 g lactose solution and subsequent measurement of breath hydrogen on both devices at 1-hour, 2-hour, and 3-hour time points. During each of these measurements, participants completed a Likert-scale assessment of gastrointestinal symptom severity (0, none; 1, mild; 2, moderate; 3, moderately severe; 4, severe; 5, very severe).

Patient Population

Adults aged 18-55 years with a self-reported history of lactose malabsorption or lactose intolerance were recruited. Exclusion criteria included history of prior gastrointestinal surgery; self-reported history of any chronic gastrointestinal disease (eg, gastroesophageal reflux disease, celiac disease, Crohn disease, ulcerative colitis, pancreatitis); self-reported history of endocrine or metabolic disease that may impact gastrointestinal or colonic function (eg, hyper/hypothyroidism, diabetes); clinically significant cardiovascular, respiratory, renal, hepatic, hematologic, neurologic, or psychiatric disease for which chronic therapy (prescription or nonprescription) is required; self-reported history of allergic reaction to any drug or drug component; antibiotic use within 28 days of lactose malabsorption testing; use of nonantibiotic prescription or over-the-counter products (dietary or digestive supplements and laxatives) within 14 days of testing; self-reported use of nicotine-containing products or chronic secondhand smoke exposure within 14 days of testing; pregnancy; any other condition which in the Investigator’s opinion may adversely affect the participant’s ability to complete the study or its measures or which may pose significant risk to the participant based on medical history or physical examination; and consumption of food after midnight on the day of testing (within 12 hours) of testing or consumption of a nonwater beverage after midnight (or less than 8 hours) prior to testing.

Devices

The Micro H2 is a hydrogen monitor that has been cleared by the FDA [10] for the diagnosis of lactose malabsorption using an automatic sensor drift detection approach, which requires gas calibration [7]. The GIMate is a portable, handheld digital hydrogen monitor with a touch screen interface that detects hydrogen using a metal-oxide sensor, which does not require calibration (Figure 1). Results from breath tests are displayed to users on the digital touch screen and stored on the device.

Figure 1. The GIMate digital hydrogen monitor.
View this figure

Outcome Measures

The primary outcome measure was positive percent agreement (PPA) in the diagnosis of lactose malabsorption of the GIMate compared to the H2 Check. Additional outcome measures included negative percent agreement (NPA) and correlation between GIMate hydrogen levels and patient self-reported gastrointestinal symptom severity. The protocol for measuring hydrogen levels was consistent with recent guidelines with a baseline measurement performed, followed by consumption of a 25 g lactose solution with subsequent hourly breath hydrogen measurements for a 3-hour period using both monitors. A positive diagnosis of lactose malabsorption was defined as a breath hydrogen level increase by 20 ppm or more from baseline at any point during the 3-hour measurement period based on guidelines [8]. A secondary outcome was the correlation between self-reported gastrointestinal symptoms and GIMate hydrogen readings.

Statistical Calculations and Analyses

The calculation for PPA was (number of individuals diagnosed with lactose malabsorption with GIMate) / (total number of individuals diagnosed with lactose malabsorption with H2 Check). The calculation for NPA was (number of individuals negative for lactose malabsorption with GIMate) / (total number of individuals negative for lactose malabsorption with H2 Check). Correlations between gastrointestinal symptoms and GIMate readings were calculated using the two-sided Spearman rank-based correlation measure of association. The P value was computed for testing for correlation estimate=0.


Population Demographics

A total of 39 individuals were screened; of these, 31 were eligible to complete the lactose challenge. Demographic characteristics for the study participants are included in Table 1. The 8 participants who were excluded from the study did not meet the eligibility criteria, including no self-reported history of lactose malabsorption or intolerance (n=1); history of prior gastrointestinal surgery (n=1); history of chronic gastrointestinal disease (n=1); clinically significant condition requiring ongoing therapy (n=1); history of allergic drug reaction (n=2); and investigator determination of a condition that would pose unnecessary risk or would adversely affect participation in the study (n=2: 1 participant had a milk allergy, and 1 participant was unable to participate during normal work hours).

Table 1. Demographics of the study participants (N=31).
CharacteristicValue
Age (years), mean (SD)33.9 (7.3)
Sex, n (%)

Female17 (55)

Male14 (45)
Race, n (%)

Asian3 (10)

Black/African-American14 (45)

White/Caucasian14 (45)
Ethnicity, n (%)

Hispanic or Latino3 (10)

Non-Hispanic or non-Latino28 (90)

Outcomes

The results for the primary outcomes regarding diagnostic performance are detailed in Table 2, demonstrating a PPA of 100% and an NPA of 100%.

The relationship between GIMate and H2 Check on a per-subject basis is depicted in Figure 2.

The relationships between GIMate and H2 Check readings for all participants across all time periods are demonstrated in Multimedia Appendix 2.

The secondary outcome of correlation between gastrointestinal symptoms and GIMate hydrogen is detailed in Multimedia Appendix 3, demonstrating an overall Spearman rank correlation of 0.82 across all time periods.

Table 2. Lactose malabsorption diagnosis by detection method (N=31).

H2 Check test results
GIMate test resultsPositiveNegative
Positive180
Negative013
Total1813
Figure 2. GIMate and H2 Check hydrogen breath measurements by participant.
View this figure

Principal Results

The primary finding of this study is the 100% positive and negative percent agreement between GIMate and H2 Check for the diagnosis of lactose malabsorption, indicating equivalent diagnostic performance between both devices. This finding was also supported by the close relationship between individual hydrogen readings on both devices for each participant (Figure 1). The correlation estimate of 0.82 between GIMate hydrogen readings and gastrointestinal symptoms indicated a strong relationship between these two variables. This finding also suggests that the rise in hydrogen levels was likely associated with lactose intolerance in our study population, which is diagnosed based on gastrointestinal symptoms in the presence of lactose malabsorption. The highest correlation between symptoms and hydrogen levels was seen at the 2-hour interval, which suggests that symptom response was greatest prior to the end of formal testing at the 3-hour mark.

Prior studies examining portable hydrogen breath testing have been conducted, including a study focused on 29 adult and pediatric patients, which required a nasal prong and syringe to obtain samples and specifically included participants with comorbid gastrointestinal conditions (eg, irritable bowel syndrome, bacterial overgrowth) [11]. Another study included 12 patients with suspected lactose intolerance using a proprietary score calculated in “arbitrary units” and did not include diagnostic criteria for lactose malabsorption [12]. This study also required the use of a phone app to display data. The H2 Check was previously evaluated in 44 patients (77% female) and compared to a composite gold standard assessment that included breath, blood, and urine testing [7]. Prior studies did not conform to the 2017 consensus guidelines on hydrogen breath testing, which provided best practice recommendations on key experimental elements such as dose of lactose, frequency of breath testing interval, and diagnostic criteria.

As a result, this study was the first to validate a portable, handheld device in diagnosing lactose malabsorption using the strongest and most current evidence-based approach. In addition, it demonstrated the first completely digital and portable measurement of breath hydrogen using clinically validated endpoints in a standalone, handheld device. This study also used the most stringent eligibility criteria compared to prior studies to minimize the risk of confounding due to the impact of comorbid gastrointestinal, surgical, and medical conditions on the production and detection of hydrogen. In addition, it is the first study to compare a novel breath hydrogen device to a previously FDA-cleared device in diagnostic performance for lactose malabsorption.

Lastly, while this was not explicitly evaluated, the compact, digital interface and substantially lower cost of the GIMate make it a potentially more convenient and safer alternative to conventional testing because it can be discarded after single-person use. These features also highlight its potential application in nonclinical settings, including home use. In this context, written instructions or integrated decision support could alert patients as to when to contact their physician based on the results. Although our study was the first to demonstrate the digital transformation of hydrogen breath testing, other studies have examined digital breath testing of other gases, including carbon dioxide [13], carbon monoxide [14], and hydrogen peroxide [15].

Limitations of this study included its relatively small sample size, although it was comparable to that in prior studies and included a more balanced and diverse representation of the population. In addition, our study was the first to examine a US-based population using a portable device. An additional limitation was the inability to directly compare our results with prior studies of portable breath hydrogen measurement, as prior studies used heterogeneous testing methods that did not conform to the most recent guidelines (two of the three studies were published prior to the establishment of these guidelines).

Conclusions

This study demonstrated that the GIMate has equivalent diagnostic performance to the H2 Check in the diagnosis of lactose malabsorption. It represents the first entirely digital approach to diagnosing lactose malabsorption with a portable, handheld device using validated clinical endpoints. These findings indicate that the GIMate is a potential viable alternative for portable, handheld detection of lactose malabsorption.

Acknowledgments

The authors acknowledge Vitaly Druker for his contributions to the statistical analysis. This study was funded by Vivante Health.

Authors' Contributions

SCM designed the intervention and wrote the manuscript. ST created training protocols for the use of equipment and provided critical review of the manuscript. SMR and MS provided critical review of the manuscript and oversaw the administration of the trial, including recruitment, allocation, scheduling, and collection and cleaning of data. SKT and SH coordinated and conducted all participant study visits. All authors reviewed and approved the manuscript.

Conflicts of Interest

SCM is an officer at Vivante Health with stock options in the company. ST has consulted for Vivante Health and has stock options in the company. SMR, SKT, SH, and MS have no conflicts to declare.

Multimedia Appendix 1

CONSORT (Consolidated Standards of Reporting Trials) diagram.

PNG File , 127 KB

Multimedia Appendix 2

Scatterplot of all GIMate and H2 Check readings.

PNG File , 59 KB

Multimedia Appendix 3

Spearman rank correlations between gastrointestinal symptoms and GIMate readings.

PNG File , 44 KB

  1. Montalto M, Curigliano V, Santoro L, Vastola M, Cammarota G, Manna R, et al. Management and treatment of lactose malabsorption. World J Gastroenterol 2006 Jan 14;12(2):187-191 [FREE Full text] [CrossRef] [Medline]
  2. Gilat T, Russo S, Gelman-Malachi E, Aldor TA. Lactase in man: a nonadaptable enzyme. Gastroenterology 1972 Jun;62(6):1125-1127. [Medline]
  3. Wang Y, Harvey CB, Hollox EJ, Phillips AD, Poulter M, Clay P, et al. The genetically programmed down-regulation of lactase in children. Gastroenterology 1998 Jun;114(6):1230-1236. [CrossRef] [Medline]
  4. Labayen I, Forga L, González A, Lenoir-Wijnkoop I, Nutr R, Martínez JA. Relationship between lactose digestion, gastrointestinal transit time and symptoms in lactose malabsorbers after dairy consumption. Aliment Pharmacol Ther 2001 Apr;15(4):543-549 [FREE Full text] [CrossRef] [Medline]
  5. Shaw AD, Davies GJ. Lactose intolerance: problems in diagnosis and treatment. J Clin Gastroenterol 1999 Apr;28(3):208-216. [CrossRef] [Medline]
  6. Jellema P, Schellevis FG, van der Windt DAWM, Kneepkens CMF, van der Horst HE. Lactose malabsorption and intolerance: a systematic review on the diagnostic value of gastrointestinal symptoms and self-reported milk intolerance. QJM 2010 Aug;103(8):555-572. [CrossRef] [Medline]
  7. Peuhkuri K, Poussa T, Korpela R. Comparison of a portable breath hydrogen analyser (Micro H2) with a Quintron MicroLyzer in measuring lactose maldigestion, and the evaluation of a Micro H2 for diagnosing hypolactasia. Scand J Clin Lab Invest 1998 May;58(3):217-224. [CrossRef] [Medline]
  8. Rezaie A, Buresi M, Lembo A, Lin H, McCallum R, Rao S, et al. Hydrogen and methane-based breath testing in gastrointestinal disorders: the North American consensus. Am J Gastroenterol 2017 May;112(5):775-784 [FREE Full text] [CrossRef] [Medline]
  9. Instructions for patients. Johns Hopkins Medicine Gastroenterology and Hepatology Breath Testing Services.   URL: https:/​/www.​hopkinsmedicine.org/​gastroenterology_hepatology/​_forms/​patient_info/​Breath_test_instructions_updated_6_5_18.​pdf [accessed 2021-08-10]
  10. 510(k) Premarket Notification. US Food and Drug Administration.   URL: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm?ID=K963376 [accessed 2021-08-10]
  11. Lee WS, Davidson GP, Moore DJ, Butler RN. Analysis of the breath hydrogen test for carbohydrate malabsorption: validation of a pocket-sized breath test analyser. J Paediatr Child Health 2000 Aug;36(4):340-342. [CrossRef] [Medline]
  12. Shrestha A, Prodhan UK, Mitchell SM, Sharma P, Barnett MPG, Milan AM, et al. Validity of a portable breath analyser (AIRE) for the assessment of lactose malabsorption. Nutrients 2019 Jul 17;11(7):1636-1636 [FREE Full text] [CrossRef] [Medline]
  13. Lorenz KA, Yeshurun S, Aziz R, Ortiz-Delatorre J, Bagley JR, Mor M, et al. A handheld metabolic device (lumen) to measure fuel utilization in healthy young adults: device validation study. Interact J Med Res 2021 May 17;10(2):e25371 [FREE Full text] [CrossRef] [Medline]
  14. Marler JD, Fujii CA, Wong KS, Galanko JA, Balbierz DJ, Utley DS. Assessment of a personal interactive carbon monoxide breath sensor in people who smoke cigarettes: single-arm cohort study. J Med Internet Res 2020 Oct 02;22(10):e22811 [FREE Full text] [CrossRef] [Medline]
  15. Neville DM, Fogg C, Brown TP, Jones TL, Lanning E, Bassett P, et al. Using the Inflammacheck device to measure the level of exhaled breath condensate hydrogen peroxide in patients with asthma and chronic obstructive pulmonary disease (the EXHALE pilot study): protocol for a cross-sectional feasibility study. JMIR Res Protoc 2018 Jan 30;7(1):e25 [FREE Full text] [CrossRef] [Medline]


FDA: US Food and Drug Administration
NPA: negative percent agreement
PPA: positive percent agreement


Edited by G Eysenbach; submitted 18.08.21; peer-reviewed by P Ray; comments to author 17.09.21; revised version received 17.09.21; accepted 20.09.21; published 18.10.21

Copyright

©Simon C Mathews, Sandy Templeton, Stephanie K Taylor, Sten Harris, Margaret Stewart, Shruti M Raja. Originally published in JMIR Formative Research (https://formative.jmir.org), 18.10.2021.

This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Formative Research, is properly cited. The complete bibliographic information, a link to the original publication on https://formative.jmir.org, as well as this copyright and license information must be included.